Page 6 - BIOFIL
P. 6

EN : All-germ point-of-use filters
                                                            DE : Endständige Keimfilter
                                                                                                       DE : Endständige Keimfilter
                                                            NL : Antibacteriële eindfilters
                                                                                                       NL : Antibacteriële eindfilters
                                                            DK : Antibakterie-endefiltre               EN : All-germ point-of-use filters
                                                                                                       DK : Antibakterie-endefiltre
                             / Point-of-use filters


                                   BACTERIAL CHALLENGE
                                                            EN : All-germ point-of-use filters         EN : All-germ point-of-use filters
                                                            DE : Endständige Keimfilter                DE : Endständige Keimfilter
                                                            NL : Antibacteriële eindfilters
                                   BIOFIL filters are approved using the ASTM F838*   Individual packet label  NL : Antibacteriële eindfilters
                                                            DK : Antibakterie-endefiltre
                                                                                                       DK : Antibakterie-endefiltre
                                   method.                                    All BIOFIL filters have a unique batch number
                                   This test confirms the effective bacterial retention power   which can be easily traced back through
                                   of the filters used for the decontamination of liquids.   the production process.
                                   The microbiological challenges carried out on
                                   BIOFIL filters in laboratories using the species
                                   Brevundimonas diminuta, Legionella pneumophila
                                   and Pseudomonas aeruginosa demonstrate
              Bacteriologically controlled water                                   30261          Qty: 1  BIOFIL  2
                                   a retention efficiency of log  for all waterborne
                                                       7
                                   micro-organisms which are larger than the filter pores
                                                                                                       FR : Filtres terminaux anti-tous germes
                                   which have a nominal porosity rating of 0.1 m    STERILE            EN : All-germ point-of-use filters
                                                            EN : All-germ point-of-use filters
                                                                                     NON
                                                                                                       DE : Endständige Keimfilter
                                                            DE : Endständige Keimfilter
                                                            NL : Antibacteriële eindfilters
                                   and absolute porosity rating of 0.2 m.      MD                      NL : Antibacteriële eindfilters
                                                                                                       DK : Antibakterie-endefiltre
                                                            DK : Antibakterie-endefiltre
                                                                                                       PL : Filtry przeciw wszystkim drobnoustrojom
                                   When subjected to this test, all BIOFIL filters delivered   18, rue du Maréchal Foch   RO : Filtre terminale împotriva germenilor
                                                                                            F-80130 Friville
              Filter        Hollow   an effluent free from bacteria, so they have a sterilising    T : +33 (0)3.22.60.22.74  ES : Filtros terminales antigérmenes de todo tipo
                                                                                                       PT : Filtros terminais anti todos os germes
              unit          fibres                                               UDI         www.delabie.com  IT: Filtri terminali antibatterici
                                   grade of 0.1 micron nominal-rated.
              Resin                                                            GTIN

                                   * Standard Test Method for Determining Retention of Membrane
                                    Filters Utilized for Liquid Filtration.   Double labelling on the filter
                                                                              All filters are supplied with 2 waterproof labels
              Section of the BIOFIL tap filter                                 which identify the product and ensure traceability
                                   LIFESPAN
                                                                              when the filter is changed.
                                                                              One is positioned on the product, the other
                                   BIOFIL anti-Legionella and all-germ filters can   can be removed and transferred to the record log,
                 2        3        be used for up to 4 months after initial installation   ensuring traceability.
                                   (depending on the model installed).
                MONTHS  MONTHS     After the time period marked on the filter (2, 3                
                                   or 4 months), DELABIE recommends changing
                      4            the filter to avoid any risk of retro-contamination
                    MONTHS         from the bacteria concentrated in the filter especially
                                   in at-risk areas.
              Maximum lifespan     Note: The amount of impurities in the water will vary
                                   between water systems. Filters will, therefore,
                                   become clogged at different rates.
                                   If the filter becomes clogged prematurely,
                                   it is necessary to change it. We recommend    COMPLIANCE
                                   pre-filtration upstream to filter out suspended solids
                                   and/or colloids, etc.                      CE marked class I Medical Devices
                                                                              BIOFIL non-sterile filters comply with the European
                                   Bacterial proliferation is significantly reduced    Regulation EU 2017/745 and benefit
                                   in an installation where the water is pre-filtered,    from the CE Mark (2022).
                                   and the lifespan of all equipment is greatly increased.
                                                                              CE marked class Is Medical Devices
              Tilted to optimise the position    RESISTANT TO DIFFERENT WATER   BIOFIL sterile filters comply with the European
              of the tap filter                                                Regulation EU 2017/745 and benefit
                                   SYSTEM TREATMENTS
                                                                              from the CE mark (2022).
                                   BIOFIL filters will withstand thermal and chemical
                                   shocks. During chemical or thermal shocks, impurities  Sterilisation complies with European standard
                                   become trapped in the filter and reduce its lifespan.  EN ISO 11137
                                   DELABIE, therefore, recommends changing the filter  After manufacture, BIOFIL sterile filters are sterilised
                                   after any treatment.                       using gamma rays. Each box has a visual marker
                                                                              showing that the filters have been sterilised.
                                   Resistance to thermal shocks:              After sterilisation, BIOFIL filters have a shelf life of
                                   Temperatures of 70°C for a cumulative period  3 years. The product expiry date is marked on the label.
                                   of 30 minutes during its lifetime.
                                                                              Certificate of sanitary compliance
              CE marked class I Medical   Resistance to chemical shocks:      All BIOFIL filters conform to the CPDW directive
              Devices for non-sterile filters   Chlorine levels for 2-month filters: 2 hours at 100ppm;
              CE marked class Is Medical   3-month filters: 3 hours at 100ppm;   and the Order of 29 May 1997, as amended,
              Devices for sterile filters                                      and the French Ministry of Health Circular
                                   and 4-month filters: 4 hours at 100ppm.
                                                                              DGS/SD7A2002 no. 571 of 25 November 2002.
                                   OPTIMISED TRACEABILITY                     WRAS
                                                                              Our BIOFIL filters are WRAS approved. WRAS
                                   To control the health risks associated with water,  Approval demonstrates that a product complies
                                   the whole process from the manufacture     with the Water Supply (Water Fittings) Regulations
                                   of the medical device components to the use    and Scottish Water Byelaws. The WRAS approval
                                   of the finished product must be completely traceable.  standard is accepted as a quality benchmark by
                                                                              water suppliers who enforce the Water Regulations
                                                                              and Byelaws.
      3
   1   2   3   4   5   6   7   8   9   10   11